PGC

Features and benefits

  • Compact 3 channel operating unit for a pressure range for passive sensors of 10-9 to 1000 mbar PGC201, 10-12 to 1000 mbar PGC202
  • Automatic switchover from PIRANI operation to Penning cold cathode operation (PGC201), UHV sensors either Bayard-Alpert measurement system IG40 BA or extractor measurement system IG40 EX (PGC202)
  • Measurement cable length up to 50 meters
  • Easy to operate
  • Two adjustable switching thresholds with a relay contact for each measurement channel
  • Logarithmic chart recorder output 0-10 V or 2-10 V
  • Wide range power supply 100 - 240 V
  • Unit of pressure selectable between mbar, Torr and Pascal
  • Compact, rugged Penning (CPG) sensor insensitive to operation at high pressures
  • Aligned and temperature compensated Pirani (PRG) sensors
  • Cost-effective replacement sensors and electrodes
  • Erro message for each channel, for example in the case of broken filament, defective sensor line or failed plasma discharge
  • Compact benchtop enclosure (1/4 19", 3 HU) made of metal for installation in front panel cut outs and 19" racks
  • RS 232 interface
  • CE mark
  • RoHS compliant

Typical applications

  • Universal pressure monitoring of high vacuum pump systems: turbomolecular, diffusion, cryogenic, ion etc.
  • Annealing, melting, brazing and hardening furnaces
  • Coating systems
  • Analytical instrumentation
  • Deployment in thermal radiation resistant and degassable systems is possible
  • Particle accelerators
Technical data Pirani / Penning Controller PGC201 Pirani / Ion Controller PGC202
Number of measurement channels 3 3

Measurement range (mbar)

Channel 1, 2 (PRG)

Channel 3 (CPG)

Channel 3

- IG40 BA

- IG40 EX

 

5 x 10-4 to 1000

10-9 to 10-2

 

-

-

 

5 x 10-4 to 1000

-

 

2 x 10-11 to 1 x 10-2

2 x 10-12 to 1 x 10-4

Unit of measurement (selectable)

mbar, Torr, Pa mbar, Torr, Pa
Measurement uncertainty    
PRG ≤20% of the measured value in the range 10-3 to 10-2 mbar (±20%) in the range 10-2 to 10-2 mbar (±15%) ≤20% of the measured value in the range 10-3 to 10-2 mbar (±20%) in the range 10-2 to 10-2 mbar (±15%)
CPG ±30% of the measured value in the range 10-8 to 10-4 mbar  
IG40 BA/EX   +/- 2% of the measured value
Measurement cable (m) up to 50 (application dependent) up to 50 (application dependent)
Display for measured values digital, 7 segment LED, 4 digit mantissa and 2 digit exponent digital, 7 segment LED, 4 digit mantissa and 2 digit exponent
Type of gas (selectable) factor adjustable factor adjustable
Operating mode switching thresholds 2 per channel single, interval-trigger 2 per channel single, interval-trigger

Adjustable switching thresholds (mbar)

PRG

5 x 10-3 to 500 5 x 10-3 to 500
CPG (mbar) 1 x 10-8 to 9.9 x 10-3  
IG40 BA (mbar)   1 x 10-8 to 5 x 10-3
IG40 EX (mbar)   1 x 10-11 to 1 x 10-11
Switching relay hysteresis 10% of the trigger value (default), freely adjustable for PRG and CPG 10% of the trigger value (default), freely adjustable for PRG and IG40 BA or EX
Relay contact load rating a.c./d.c., max. 30 V/1 A a.c./d.c., max. 30 V/1 A

Chart recorder output (default)

PRG

0 to 10 V, log. divisions linear: 3 decades, approximately 10.5 V in case of a failure, logarithmic: (1 x 10-3 mbar), 1.67 V/decade 0 to 10 V, log. divisions linear: 3 decades, approximately 10.5 V in case of a failure,
logarithmic: (1 x 10-3 mbar), 1.67 V/decade

CPG

logarithmic: (1 x 10-9 mbar), 1.43 V/ decade  

IG40 BA or EX

  logarithmic (1 x 10-12 mbar), 1,00 V/decade

Interface

RS 232, RS 485 RS 232 C, RS 485

Mains connection 50/60 Hz (V a.c.)

100 - 240 100 - 240

Power consumption (W)

<10 <60

Storage temperature range (°C)

-20 to +60 -20 to +60

Nominal temperature range (°C)

+5 to +50 +5 to +50

Max. rel. humidity

80 80

Weight (kg)

1,4 1,4

Dimension (W x H x D) (mm)

106,4 x 128,4 x 174,5 106,4 x 128,4 x 184,5

Installation depth (mm)

approx. 220 approx. 220

Protection class (IP)