Technical data | Pirani / Penning Controller PGC201 | Pirani / Ion Controller PGC202 |
Number of measurement channels | 3 | 3 |
Measurement range (mbar) Channel 1, 2 (PRG) Channel 3 (CPG) Channel 3 - IG40 BA - IG40 EX |
5 x 10-4 to 1000 10-9 to 10-2
- - |
5 x 10-4 to 1000 -
2 x 10-11 to 1 x 10-2 2 x 10-12 to 1 x 10-4 |
Unit of measurement (selectable) |
mbar, Torr, Pa | mbar, Torr, Pa |
Measurement uncertainty | ||
PRG | ≤20% of the measured value in the range 10-3 to 10-2 mbar (±20%) in the range 10-2 to 10-2 mbar (±15%) | ≤20% of the measured value in the range 10-3 to 10-2 mbar (±20%) in the range 10-2 to 10-2 mbar (±15%) |
CPG | ±30% of the measured value in the range 10-8 to 10-4 mbar | |
IG40 BA/EX | +/- 2% of the measured value | |
Measurement cable (m) | up to 50 (application dependent) | up to 50 (application dependent) |
Display for measured values | digital, 7 segment LED, 4 digit mantissa and 2 digit exponent | digital, 7 segment LED, 4 digit mantissa and 2 digit exponent |
Type of gas (selectable) | factor adjustable | factor adjustable |
Operating mode switching thresholds | 2 per channel single, interval-trigger | 2 per channel single, interval-trigger |
Adjustable switching thresholds (mbar) PRG |
5 x 10-3 to 500 | 5 x 10-3 to 500 |
CPG (mbar) | 1 x 10-8 to 9.9 x 10-3 | |
IG40 BA (mbar) | 1 x 10-8 to 5 x 10-3 | |
IG40 EX (mbar) | 1 x 10-11 to 1 x 10-11 | |
Switching relay hysteresis | 10% of the trigger value (default), freely adjustable for PRG and CPG | 10% of the trigger value (default), freely adjustable for PRG and IG40 BA or EX |
Relay contact load rating | a.c./d.c., max. 30 V/1 A | a.c./d.c., max. 30 V/1 A |
Chart recorder output (default) PRG |
0 to 10 V, log. divisions linear: 3 decades, approximately 10.5 V in case of a failure, logarithmic: (1 x 10-3 mbar), 1.67 V/decade | 0 to 10 V, log. divisions linear: 3 decades, approximately 10.5 V in case of a failure, logarithmic: (1 x 10-3 mbar), 1.67 V/decade |
CPG |
logarithmic: (1 x 10-9 mbar), 1.43 V/ decade | |
IG40 BA or EX |
logarithmic (1 x 10-12 mbar), 1,00 V/decade | |
Interface |
RS 232, RS 485 | RS 232 C, RS 485 |
Mains connection 50/60 Hz (V a.c.) |
100 - 240 | 100 - 240 |
Power consumption (W) |
<10 | <60 |
Storage temperature range (°C) |
-20 to +60 | -20 to +60 |
Nominal temperature range (°C) |
+5 to +50 | +5 to +50 |
Max. rel. humidity |
80 | 80 |
Weight (kg) |
1,4 | 1,4 |
Dimension (W x H x D) (mm) |
106,4 x 128,4 x 174,5 | 106,4 x 128,4 x 184,5 |
Installation depth (mm) |
approx. 220 | approx. 220 |
Protection class (IP) |
40 | 40 |