Technical data | |
Pumping speed | |
nitrogen | 2000 l s-1 |
hydrogen | 3000 l s-1 |
Min backing pump displacement* | 40 m3h-1 |
Recommended backing pump | E2M40 |
Recommended fluid | Santovac® 5 |
Fluid charge (dry) | 500 ml |
Inlet connection compatible with | ISO250 |
Backing connection | NW40 |
Cooling water connection | 10 mm compression fittings |
Heater power | 2.25 kW |
Min cooling-water flow at 20°C | 180 l h-1 |
Pneumatic connections† | 6mm coupling x ¼ BSP male stud |
Pneumatic actuating pressure | |
Minimum | 2.4bar / 35psi |
Maximum | 6.9bar / 100psi |
Weight | |
250/2000M | 59kg |